In the case where Overjet times out, that means Sidexis is not talking to the Pano server. If Sidexis will not launch, do the following.


Delete the files found in on the Pano server

D:\PDATA\Viewsets
D:\PDATA\RawDataEraser 


Restart the services, Sirona AcquisitionServer and Sidexis Service.